Focus calibration is a calibration of the drum and should be carried out
when the scanner is first installed and each time a different drum is
mounted.
Technical data
-----
Tools and materials required
Mac or PC (both if possible)
Photoshop plug-in 2.1.7 or later versions
Auto focus adjustment target
5 • ADJUSTMENTS AND TESTING
Focus Calibration
Edition 1 January 1996 Service Manual • ScanMate 4000/5000
5•4a
Caution! Turn off power • Before connecting or disconnecting cables.
• Before carrying out service operations.