Service Manual • ScanMate 4000/5000 Edition 1 January 1996
8•4
8 • TROUBLE SHOOTING
Problem Possible Cause Solution
General
Monitor program is not 1. Dip switch setting incorrect. Set Dip switch to 8.
running
Calibration fails 1. Aperture plate out of adjust- Calibrate aperture (see
ment. section 3•8).
No ready light. 1. Defective park position Replace switch
switch. Contact ScanView support.
No lights on front panel. 1. Cable faults. Replace cable(s).
2. Defective power supply. Replace power supply.
Contact ScanView support.
Nothing is working. 1. Fuse in the scanner blown. Change the fuse.
Contact ScanView support.
SCSI. 1. Dip switch setting incorrect. Set Dip switch to the SCSI
number that is required.
Contact ScanView support.