3 • SERVICE PROGRAM
Edition 1 • January 1996 Service Manual • ScanMate 4000/5000
3•11
Status received : HBCR-161. Version 15
Found barcode chip
Manually rotate the drum slightly and the following will be shown:
Got string : xxxxxxxxxxxxx
For example, 000002889405 where 0288 is the drum serial number.
000002889405 where 940 is the drum diameter of 99.4
000002889405 where 5 is the bar code check number.
Show scanner options (*)
This command facility is used to show the gain, aperture and focus con-
stants. It can also be used for sending a logfile to ScanView’s support
team.
Select * to enter the Show scanner options menu and the all the constants
will be shown.
Note: The values shown are only examples.
>> ApertureValue 0 : 20.100000
ApertureValue 1 : 4.920000
ApertureValue 2 : 4.180000
ApertureValue 3 : 3.750000
ApertureValue 4 : 2.210000
ApertureValue 5 : 2.660000
Aperture 1, PMTTransGain PMTReflectGain : 51 60
Red TransGain, ReflectGain : 110 131
Green TransGain, ReflectGain : 106 130
Blue TransGain, ReflectGain : 100 103
Aperture 2, PMTTransGain PMTReflectGain : 159 60
Red TransGain, ReflectGain : 115 131
Green TransGain, ReflectGain : 107 130
Blue TransGain, ReflectGain : 107 103
Aperture 3, PMTTransGain PMTReflectGain : 107 60
Red TransGain, ReflectGain : 103 131
Green TransGain, ReflectGain : 110 130
Blue TransGain, ReflectGain : 104 103
Aperture 4, PMTTransGain PMTReflectGain : 135 60
Red TransGain, ReflectGain : 91 131
Green TransGain, ReflectGain : 110 130
Blue TransGain, ReflectGain : 104 103
Aperture 5, PMTTransGain PMTReflectGain : 135 93
Red TransGain, ReflectGain : 91 125
Green TransGain, ReflectGain : 110 130
Blue TransGain, ReflectGain : 104 103