EasyManua.ls Logo

SCANVIEW ScanMate 5000 - Focus Sharpness

Default Icon
176 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Function
The testing of focus sharpness checks that maximum sharpness is ob-
tained to ensure the best possible focus.
Technical data
Focus: < 6 micron
Tools and equipment required
Photoshop plug-in 2.1.7 or latest version
UGRA test target to measure the focus value.
5 • ADJUSTMENTS AND TESTING
Focus Sharpness
Edition 1 January 1996 Service Manual • ScanMate 4000/5000
5•11a
Caution! Turn off power • Before connecting or disconnecting cables.
• Before carrying out service operations.
Ensure the UGRA test target is free of dust otherwise false indica-
tions are likely.

Table of Contents