Function
The zero point vertical adjustment ensures that the starting point of the
scanner is set correctly.
Technical data
-----
Tools and equipment required
2 mm allen key for back cover mounting screws
2.5 mm allen key for DC motor covers and adjustment screws
2.5 mm allen key (long) for lamp cover mounting screws
3 mm allen key for right hand end plate mounting screw
3 mm allen key (T-handle) for right hand drum cover mounting screws
Oscilloscope with 2 channels
Extension board (part of the service kit)
5 • ADJUSTMENTS AND TESTING
Zero Point Vertical
Edition 1 • January 1996 Service Manual • ScanMate 4000/5000
5•8a
Caution! ☛ Turn off power • Before connecting or disconnecting cables.
• Before carrying out service operations.