Standard Equipment Required
Voltmeter
20 MHz (minimum) oscilloscope, 2 channels, probes
Service Kit
1 C-spanner for light tube adjustment
1 ScanView liquid drum polish, including one cloth
1 ScanView anti-static cloth
2 ScanView grease for spindle and shaft (2 x 3 g. tubes)
1 ScanView Monitor program
1 3. 5” 1.44 MB DOS 5.0 disk with ScanView service program ver. 1.0.2
1 2 m. RS232 cable for PC to scanner communication
1 2 m. RS232 cable for Mac to scanner communication
1 9 to 25 pin converter
2 ScanView focus target, (focus adjust original)
1 UGRA offset test target
1 ScanView 4000/5000 service manual
1 3.0 mm T-handle allen key
1 Extender board
Error Reporting
The aim of the Error Report is to give ScanView A/S a standard error
reporting medium where all the necessary information is given to Scan-
View’s Support Department to enable the assessment of defects to be
made in a rational and effective way.
A master copy of this error report is shown on page 1•13 and should be
used to take all the necessary copies.
Of particular importance, is the allocation of a CRN (Confirmation of
Return Number). This number is obtained by contacting ScanView (Sup-
port departments hot-line, telefaxing or using the BBS) and must be
included on all error reports.
Exchange parts must be returned to ScanView within 3 weeks of the
invoice date.
Service Manual • ScanMate 4000/5000 Edition 1 January 1996
1•10
1 • GENERAL INFORMATION