How to replace
Removal
1 Remove the following covers:
Left hand end plate E and left hand drum cover F, (section 4•2 steps
5 and 6).
2 In order to give access to the spindle step motor, move the drum to
the zero position, turn on power and access the service program,
(section 3•5).
3 Select 1, (Move spindle) and Z, (Set drum in zero position). Turn off
power.
4 Disconnect the connector to the step motor.
5 Manually rotate the spindle until the head of the 4 mm dome head
allen screw A securing the step motor to the spindle is accessible
and remove the screw and lock washer.
6 Remove the four 4 mm allen screws B retaining the step motor to
its mounting bracket and lift the motor away.
Assembly
7 Mount a new step motor and refit the four 4 mm allen screws.
8 Align the holes in the step motor shaft and spindle and refit the 4
mm dome head allen screw and new lock washer.
9 Reconnect the connector to the step motor.
10 Carry out the adjustment and test operations specified below.
11 If no further replacement operations are required, refit the covers
removed in step 1 in accordance with section 4•2.
Adjustments required
White calibration in transmission and reflection (section 5•5)
Register error (section 5•12).
Service Manual • ScanMate 4000/5000 Edition 1 January 1996
Spindle Step Motor
4•14b