Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 *
http://www.schleich.com * info@schleich.com 13
1.3 Range of application
The safety and function tester GLP2-i/e is designed for testing voltage-free 2-poles only (unless it is a 3-
phase model or a model made to specification).
Any usage other than this is regarded as not appropriate.
For any damage resulting from inappropriate usage, the manufacturer/supplier can not be hold liable; the
user bears all risks.
Before the tester is switched on for the first time or after it has been transported, it needs to adapt to the
ambient temperature first. Please unpack the device and put it on a table. You should wait at least 30
minutes before switching it on.
The device must not be exposed to the condensation of water.
Attention!
We are pointing out that for the daily check of the testing device regarding its correct
functioning and accuracy, it is necessary to use a calibrated black box (test dummy). This black
box is connected to the test leads of the testing device and it is tested, whether the data of the
black box are the same as the measured values. In case the testing device has various
measuring ranges, it is necessary that for each measuring range the respective black box is
used.
In order to make sure that the comparison data are correct, the black box must be checked
annually.
For testing with the black-box, the testing device has a suitable program. Right after switching
on the testing device, this program asks the person who is doing the test to do the self-test with
the black-box. Only after the self-test has been completed correctly, the device enables itself for
further tests.
Please note: In the SetUp of the testing device it is possible for the user to deactivate the self-test.
However, doing this is not in accordance with the intended purpose of the device. If the self-test is
deactivated, there will also be a message on the display.
Attention!
We should like to point out: Every year and after every modification (repair), which could influence the
measuring equipment, the testing device must be calibrated and, if necessary, adjusted. The
supplier/manufacturer will test the device regarding its accuracy and confirm the calibration in a certificate.
Attention!
For test methods, their measuring ranges and accuracies, please refer to the enclosed calibration certificate.