Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 *
http://www.schleich.com * info@schleich.com 46
5.5.4 General data
General data
Self-test..............: At power on
Start with probe.......: on
Start saf. Switch......: off
Safety ISO.............: off
Contin. function.......: off
no go buzzer...........: off
FCT with +-(%) tol.....: off
Offset PE test.........:
0.00
Repeat if no go........: 3
Shrt.circ.resis........:
25
VIS switch for go......: Probe switch
VIS switch for no go...: Start switch
With serial number.....: off
With name of user......: off
With order inputs......: off
Test step display......: Set values
Contrast control.......: 175
Enter & ?
- +
Self-test (Blackbox test)
Please enter, when the self-test is to be performed.
At power on: The self-test is performed when the device is switched on
At date change: The self-test is performed when the date is changed (if a clock is installed)
Start with probe
Please enter, whether the test process can be started also with the switch inside the test probe.
Start saf. switch
Please enter, whether the test process can be started also by closing the safety switch.
The device has two safety switches: HV contact and LV contact. You can select, which contact is to be used
for starting the test.
Safety ISO (only for devices with insulation test)
Please enter, whether the safety switch LV is active also during the insulation test.
Safety test cover (only for special test covers or dual covers)
Please enter, whether the test cover is to stay shut during the test.
Start contacting (only for special contactings)
The test is started by closing the automatic contacting.
order input