Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 *
http://www.schleich.com * info@schleich.com 9
1.2.5 Leakage current test
Leakage current: max. 30mA
Resolution: 0.01mA
Accuracy: 0.2% of measuring range
Testing according to EN60990 requires 3 measuring circuits for the evaluation.
1: unweighted leakage current (EN60990 (5) Measuring Equipment Fig.3)
2: weighted according to perception and reaction (EN60990 (5) Measuring Equipment Fig.4)
3: weighted after release (EN60990 (5) Measuring Equipment Fig.5)
Touch-current frequency: max. 500Hz
Evaluation: r.m.s value (unweighted) or peak value
Warm-up time: 0 seconds ... 24 hours
Max. current at "warm leakage current test": like function test
IMPORTANT
In case the test object needs more than 5A, it is necessary to order the increased current additionally.
TESTING STANDARDS
EN60990 / VDE 0106
EN60335-1 / VDE 0700-1 Measuring at household devices
EN60950 / VDE 0805 Measuring at office devices (communication devices)
EN60598-1 / VDE 0711-1 Measuring at lamp
EN60990 (5) Measuring Equipment Fig.3
EN60990 (5) Measuring Equipment Fig.4
5) Measuring Equipment Fig.5