Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 *
http://www.schleich.com * info@schleich.com 60
5.6 Single test
The tester GLP2-i allows to perform single tests of the test methods integrated into the device. The test
methods are indicated in the menu “Single test”. The method is selected by pressing the respective button.
The advantage of the single test is that the complete display is used for indicating the measured values.
The single test is well suited for specific trouble shooting at your test objects, because the individual
measuring values are directly visible and the test doesn’t need to be restarted every time.
For starting a single test, you only need to activate the start switch at the device or the foot switch. For
switching off, you activate the switch again or press the ESC button on the display.
Single test
Protective earth High voltage
Insulation Continuity
Function Resistance
ESC
Example of a single test (insulation test):
Single test: Insulation test
R= >500M
I=0,0000mA
U=997,4V
ESC - N - PE1 + - 1000 V +
Control of Start and Stop of the single test via the normal start switch on the front panel.
Selection of a test method for the
single test (according to the
available test methods).
Actual value of the insulation resistance
Exit single test: insulation test
Actual value of the insulation cur
Actual value of the insulation test
voltage
Connection point of the test object /
test connections; change connections
with + and - .
Test voltage; change voltage with
+ and - .