Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 *
http://www.schleich.com * info@schleich.com 29
This message must be acknowledged by touching the display. Now, the testers indicates a message about
using the Blackbox. Standard DIN ISO 10012 requires the user of a tester to check the device in regular
intervals with a blackbox. This message will only appear, if the self-test via blackbox is deactivated. The
message must be acknowledged by touching the display.
Test without self-test!
Before the test is started, a
self-test can be done by connect-
ing a blackbox. This serves to
check the function of the testing
device. For this purpose, the
database must contain a test
program defined as BLACKBOX.
Now, the menu for testing will be indicated.
No. 100W Lamp St.vl At.vl Res
1 PE (1)
0,2
2 PE (1)
0,2
3 ISO(1)
2,0M
4 ISO(1)
2,0M
5 FCT(1) 0,45A
6
7
8
9
10
Input noG 0 0 sec go 0