Item Condition
Confirmed
1-1 Temperature
Guaranteed within specification
Guaranteed operation
18 to 28
M
C (constant)
15 to 35
M
C
1-2 Humidity 40% to 70%
1-3 Installation location
Space
Strength
GC/MS unit approx. 75kg
PC unit approx. 20kg *
Space at rear
Space at left
* Weight may vary according to the
model of PC, monitor, and printer.
2000mm (W) x 700mm (D)
Able to support weight
(approx. 110kg.)
300mm min.
400mm min.
1.
Installation
room
1-4 Others Few harmful elements, such
as dust, vibrations,
electromagnetic fields, and
corrosive gas.
2-1 Voltage fluctuations
115V instrument
Guaranteed within specification
Guaranteed operation
230V instrument
Guaranteed within specification
Guaranteed operation
100V
±
5%
100V
±
10%
100V
±
5%
100V
±
10%
2-2 Frequency fluctuations 50/60 Hz
±0.5 Hz within
2-3 Terminals
* 115V instrument
GC unit (Fig. 1)
MS unit (Fig. 1)
* 230V instrument
GC unit (Fig. 1)
MS unit (Fig. 1)
Separate power supplies are required for
the PC, monitor, and printer.
Cable length
Capacity
2m
1800VA
2.5m
1000VA
2m
2600VA
2.5m
1000VA
2. Power
supply
2-4 Grounding resistance
100
Ω
max.
3. Gas 3-1 Helium
Purity
Regulator
Pipe between regulator and GC
99.995% min
PPR-He, or equivalent
Carrier gas pipe
201-48067 2.5m
201-48067-05 5m
201-48067-10 10m
201-48067-15 15m