200-33168
Seal Teflon
As required Consumable
036-15550-12
Oring,AS568A-012 4D
As required Consumable
225-09344
Sample vacuum bottle
As required Consumable
225-04209
cap
As required Consumable
225-10627-91
DI PROBE ASSY
As required Consumable
DI
225-10618-91
REPELLER2,CI+(C)
IN DI mode ,CI analysis
As required Consumable
225-09493-03
STANDARD SAMPLE (PFTBA) 5g
As required Consumable
STANDAR
DSAMPLE
225-09493-01
STANDARD SAMPLE (TRIAZINE) 0.5g
As required Consumable
Consumable
VACUM
SYSTEM
017-30163-11
ROTARY PUMP OIL ULTRAGRADE15 1L 0.28L per oil replacement 1L/year (#5)
Consumable
085-35124-03
SCOTCHBRITE (for finishing electrodes) set of 20 sheets
20 sheets/year
225-10194-91
JIG, IS
Service part
225-01453-94
COLUMN MOUNTING JIG (I/F)
Tools and
jigs
225-01453-95
COLUMN MOUNTING JIG (INJ)
#1 Replace after approximately 80 injections.
#2 If carrier gas leaks or when glass insert is replaced.
#3
Replace periodically, normally once every 6 months. However, replace more frequently under
certain measurement conditions or after many analyses of samples with a high boiling-point or
samples that are solid at room temperature.
(See GC System User Guide for details.)
#4 Use when column is replaced.
#5 Ideally, replace oil once every three months (3000 hours operating time). Insufficient vacuum can cause various problems in the
mass spectrometer system. The oil must be replaced appropriately.