225-11075-91
RF PS ASSY Service part
225-11060-91
PCB ASSY, RF DIFF-A Service part
RF PS
225-10296-91
FAN ASSY Service part
MAIN CTRL
225-11130-91
PCB ASSY, MAIN CTRL-A Service part
CPU PCB
225-11000-91
PCB ASSY, CPU-A Service part
IS CTRL
225-11225-91
IS CTRL-A ASSY Service part
DC POWER
225-11250-91
DC POWER-A Service part
IG CTRL
225-11280-91
PCB ASSY, IG CTRL-A Service part
PRE AMP
225-11030-91
PCB ASSY, PRE AMP-A Service part
225-11310-91
PCB ASSY, DI CTRL-A DI IPTION Service part
225-11300-91
PCB ASSY,DI SOCK-A DI IPTION Service part
225-11325-91
RELAY BOX 100V DI IPTION(100/115V) Service part
DI
225-11325-38
RELAY BOX 230V DI IPTION(220-240V) Service part
225-10576-91
CI GAS CONTROLLER CI/NCI OPTION Service part
CI/NCI
225-10603-91
ELBOW,KF16-#15,CI CI/NCI OPTION Service part
071-60814-05
CORD, KP-4819D+KS31A (220 to 240V) Service part
071-60815-04
CORD SET, KP300C+KS16* (100 to 115V) Service part
070-60311-02
CABLE, AXJ231122 (IEEE-1394) Service part
070-60311-03
CABLE, AXJ231132 (4.5m) (IEEE-1394) Service part
CABLE
088-50906-11
CABLE, KRS-403XF1K (RS-232C) Service part
#1
Once every 3 years is sufficient, unless some unforeseen accident occurs.
This is an expensive part and having one available can avoid sudden expenditure. However, storage over a long period
is not recommended, as the detector deteriorates if unused for long periods.
#2 Overhaul required every 25,000 hours (approx. 2.8 years).
#3 Overhaul required every 15,000 hours (approx. 1.7 years).

Table of Contents