Day 1
1. Checking Installation Conditions
Confirm the installation conditions, according to the Installation Completion
Checksheet (ZEAV-0416).
Check the power supply.
1)power requirements
Frequency requirements
MS power supply: 115V±5%, 1000VA
 230v±5%,1000VA
GC power supply: 115V±5%, 1800VA
230V±5%,2600VA
230V±5%, 2600VA for high-power oven specification.
Not including GC options.
In addition,power for the pc,monitor,and printer must be provided.
*There may be instance in which the current capacity will increase as options are
attached.
2) Ground : 100Ω or less
3)Voltage Fluctuation Range
Specified power supply voltage range:±5%
Operating power supply voltage range:±10%
CAUTION: USE the PC,monitor,and printer that conform to the law and regulation in
respective
NOTE: Be sure sure that power is supplied from a apower sourse with a ground fault
breaker in
NOTE: Voltage fluctuations are established so that the instrument can operate at full
capacity within the range of ±10,including any transient noise in the AC line.
Note,however,that the voltage fluctuation range that guarantees the performance
specifications should be used within ±5%.The degree of frequency stability of
power source should be within ±0.5%.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GCMS-QP2010 and is the answer not in the manual?

Shimadzu GCMS-QP2010 Specifications

General IconGeneral
BrandShimadzu
ModelGCMS-QP2010
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals