V
CP3
should be 1 x conversion dynode setting (kV). Any abnormality in this voltage is
probably due to incorrect D/A voltage from the CPU PCB.
Check that when CD HV is ON, V
CP3
=V
CP1
; when CD HV is OFF, V
CP1
0V.
If the V
CP1
voltage is generated whether CD HV is ON or OFF, this suggests a defective
controller operational amplifier. Replace the MAIN CTRL PCB.
If V
CP1
= 0V when CD HV is ON, the problem is probably in the high-voltage power supply
assy, or in the connector cables.