Conversion dinode voltage
Measurement point: J12 [CD]
Inside resistance: 33M
Detection device lens voltage
Measurement point
Field through Electrode
Detection device lens voltage=0.084×conversion dinode voltage
Conversion dinode voltage, becomes big 220V each to every measure current 1uA.
When some extent is exceeded it becomes a high pressure power supply error.
Please pay attention to the input resistance of ameasuring apparatus.