EasyManua.ls Logo

Shinko PCD-33A - PID Auto-Tuning

Shinko PCD-33A
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42
[Auxiliary function setting mode 1]
Press the key 3 times.
Sensor correction setting item is selected.
(Press 3 times)
Set the sensor correction value with
or key.
Setting range: -100.0 to 100.0 ( )
, For DC input, -1000 to 1000 (The placement of
the decimal point follows the selection)
Press the key.
The controller reverts to the program standby
mode or program control run mode
[Program standby mode or program control run mode]
The value on the PV display is the one corrected
by the sensor.
(e.g.) If actual temperature is 200 , and sensor
correction value is 2.0 , PV becomes 202 .
7.7 PID auto-tuning.
In order to decide each value of P, I, D and ARW automatically, this system
makes the controlled objects temperature fluctuate.
Notice
Perform the PID auto-tuning during the trial run.
During the PID auto-tuning, none of the setting items can be set.
If PID auto-tuning is performed, PID auto-tuning starts at the same time as performance,
and step time and time signal time do not progress until the auto-tuning ends.
After the PID auto-tuning ends, the remaining step time is performed.
If power failure occurs during the PID auto-tuning, the tuning stops.
(Fig. 7.7-1)
PTN
PV
STEP
SV
Selected value
PTN
PV
STEP
SV
PTN
PV
STEP
SV
Set value
PTN
PV
Actual temp.
STEP
SV
PID auto-tuning
is performing
Delayed time due
to PID auto-tuning
: Pattern before the PID auto-tuning is performed
: Pattern after the PID auto-tuning is performed

Table of Contents