EasyManua.ls Logo

Shinko PCD-33A - Troubleshooting; Troubleshooting by Indication; Troubleshooting by Keypad Operation

Shinko PCD-33A
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
56
<Indication>
Problem Presumed cause and solution
The PV display is
blinking [ ].
Thermocouple or RTD is burnt out.
[Thermocouple]
If the input terminal of the instrument is shorted,
and if nearby room temperature is indicated,
the instrument is probably normal and the sensor
may be burnt out.
[RTD]
If approx. 100 of resistance is connected to the
input terminals between A-B of the instrument and
between B-B is shorted, and if nearby 0 (32 ) is
indicated, the instrument is probably normal and the
sensor may be burnt out.
Lead wire of thermocouple or RTD is not securely
mounted to the instrument terminals.
The PV display is
blinking [ ].
Polarity of thermocouple or compensating lead wire
is reversed.
Codes (A, B, B) of RTD do not agree with the
instrument terminals.
If indication
of PV display
is abnormal or
unstable.
Designation of the sensor input is improper.
Temperature unit ( or ) is mistaken.
Set the sensor type which is the same as users and
the temperature unit from the Input type selection
[ ]. (p.16)
Sensor correction value is unsuitable.
Set the value properly from the Sensor correction
setting [ ] in Auxiliary function setting mode 1.
(pp.31, 41, 42)
AC may be leaking into thermocouple or the RTD
circuit.
There may be equipment nearby producing an
inductive fault or noise near the controller.
PV display indicates
[ ].
Internal memory is defective.
Make inquiries to Shinko Technos or our dealers.
<Keypad operation>
Problem Presumed cause and solution
It is impossible to
select program pattern
number.
The program control is running.
Change the mode.
The program pattern number can be selected in the
program standby mode.
If settings or
selections are
impossible.
Setting value lock [ ] is designated from the
Setting value lock selection in Auxiliary function setting
mode 1.
Release the lock designation and set to [ ].
(p.31)
PID auto-tuning is performing.
Cancel the tuning if necessary. (p.28)

Table of Contents

Related product manuals