EasyManua.ls Logo

SICK DUSTHUNTER T - Page 182

SICK DUSTHUNTER T
552 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance
GMS800 · Operating Instructions · 8013025 V1.2 · © SICK MAIHAK GmbH 55
Subject to change without notice
Fig. 10 Simple test method for leak tightness check (example)
Test procedure
1 Put the gas analyzer out of operation. Separate the gas inlet and outlet of the gas ana-
lyzer from the installation (when fitted).
2 Connect the gas analyzer gas inlet to the washing bottle gas outlet.
3 Close the gas analyzer gas outlet off gas-tight, e.g. with a sealing plug.
4 Also close off all the other internal gas path connections (as available) in the same man-
ner.
5 Check: The valve on the pressure reducer gas outlet must be closed off. Now open the
main valve of the compressed gas cylinder.
6 Adjust the pressure reducer so that the output pressure (secondary pressure) is
150 kPa (1.5 bar).
7 Connect the pressure reducer gas outlet to the washing bottle gas inlet.
8 Open the pressure reducer valve slowly (avoid sudden pressure increase).
9 Wait until the pressure ratio is constant (several seconds).
10 Observe the washing bottle: Count the number of air bubbles rising for 1 minute.
The gas path is classified as tight when less than 60 air bubbles rise per minute.
11 To terminate the test procedure:
Close off the valve on the pressure reducer gas outlet.
To release the gas pressure: Carefully loosen the connection hose on the washing
bottle gas outlet.
Reconnect the gas connection on the gas analyzer as for operating mode - pay
careful attention to gas leak tightness.
Gas analyzer
When the gas analyzer has several separate internal gas paths:
b
Perform this procedure individually for each gas path.

Table of Contents

Related product manuals