EasyManua.ls Logo

SICK DUSTHUNTER T

SICK DUSTHUNTER T
552 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Instructions GM35
Probe Model
Troubleshooting
8009389/07-2006 © SICK MAIHAK GmbH • Germany · All rights reserved 83
9.3.1 Troubleshooting the evaluation unit
9.3.2 Error messages for the GM35 SR unit
The following error messages, which can be displayed on the evaluation unit, refer to the
GM35 SR unit and the purge air attachments.
Error description/message Component/possible cause Remedy
Evaluation unit not respond-
ing
Evaluation unit:
Power supply to evaluation unit
defective
Check power supply to all system components:
If necessary, connect on-site power supply
If necessary, check/reconnect the terminals on the
system components
Evaluation unit:
Incorrect operating voltage
Check the operating voltage set on the evaluation
unit:
If necessary, change setting
Evaluation unit:
Fuse defective
Check the fuse in the evaluation unit:
If necessary, change fuse
Evaluation unit:
No defect localized
Disconnect all system components from the power
supply and reconnect them one after the other
Check the CAN bus cable from the evaluation unit to the
SR unit or terminal box
Evaluation unit:
Error occurs again
Replace the component last connected; contact the
Service department
Evaluation unit:
24V/5V supply defective
Check 24V/5V supply,
Replace evaluation unit or electronic board module;
contact Service department
Evaluation unit:
Inconsistent data detected in
parameter memory
Press Enter to restart the system; the default
parameter settings are then active;
If necessary, reconfigure the parameters
If the same error message occurs again, replace the
evaluation unit and contact the Service department
Error message Component/possible cause Remedy
Air purge low SLV The volumetric flow undershoots the
set threshold.
Check the purge air supply (fan, hoses). If nec-
essary, replace the filter on the purge air fan
(see chapter 8.3.2, page 70.
CO Ampl. max CO instrument amplifier is above the
control range
Check alignment (see page 72).
Clean optical boundary surface (see page 47/
page 68).
Contact the Service department.
CO com. CO
2
/H
2
O module not connected to
CO module
Check whether the cable connection is secure.
Secure/repair it inside the GM35 if necessary.
If fault cannot be corrected: contact Service
department.

Table of Contents

Related product manuals