FLOWSIC100 Operating Instructions
Gas Velocity Monitor
Maintenance
130 © SICK MAIHAK GmbH · Germany · All rights reserved 8012513/03-2008
5.5.2 Replacing the filter insert
Have a new filter insert (2) at the ready.
Loosen the hose clamp (6) on the purge-air hose (7) and remove the hose. Secure the
hose at a clean location.
ATTENTION
Important
Place the end of the hose in a safe place so that impurities cannot be sucked in (this will
cause irreparable damage to the fan). During this time, the purge-air entering the purge-
air glands is unfiltered.
Remove any dust from the outside of the filter housing (1).
Press the two quick-release locks (4) on the filter housing cover (3) to remove it.
Remove the filter insert (2) by twisting it counterclockwise.
Clean the insert of filter housing and housing cover with a cloth and brush.
ATTENTION
Important
Only use a cloth soaked in water to wet-clean the parts. Make sure that the parts are
dried well afterwards.
Spare part:
Filter insert Micro-Top
element C11 100,
Part No. 5306091
Insert the new filter insert by twisting it clockwise.
Mount the cover on the filter housing, ensuring that it is aligned correctly with the hous-
ing, and snap the quick-release locks into position.
Connect the purge-air hose to the filter outlet again using the hose clamp.
Fig. 5.3: Replacing the filter insert
9
1234
8
7
65
1 Filter housing 6 Hose clamp
2 Filter insert 7 Purge-air hose
3 Filter housing cover 8 Hose clamp
4 Quick-release lock 9 Low-pressure monitor
5 Air intake gland