UMSITRPDS3-1 Calibration And Maintenance
November 2005
7-5
7.3 TROUBLESHOOTING
This section provides guidance and procedures to assist in identifying and correcting a malfunctioning
transmitter. Section 7.0 lists needed tools and equipment.
It is recommended that all documentation associated with the transmitter, including piping and loop
wiring diagrams and configuration documentation, be obtained and made available to maintenance
personnel to facilitate troubleshooting.
The most common symptom of a malfunctioning transmitter is incorrect, erratic, or no output. A
malfunction can affect the transmitter’s analog output (4-20 mA) or its digital (HART) output.
Furthermore, a malfunction can be the result of external forces and not a transmitter fault at all. Section
7.3.1 discusses troubleshooting techniques for the analog output. Section 7.3.2 discusses troubleshooting
techniques for the digital output.
General transmitter troubleshooting steps:
Record all failure symptoms. Record the values and messages on the transmitter display. The display
is discussed in Section 6 On-Line Calibration and Operation.
Confirm the values and messages recorded above by connecting a HART Communicator (or DMM)
and recording the values and messages shown by the Communicator. Navigate to those displays that
show active data. Refer to Appendix A Online Configuration Maps to roam the Communicator’s
menus.
Test the suspect unit by performing a transmitter self test or transmitter reset and again note the
values and messages on the transmitter display and on the Communicator.
Apply a known pressure (or level) and note all values and readings on the transmitter display and
Communicator. See Section 5 Post Installation Tests for sample procedures.
Check impulse piping, loop wiring, power supply, and related connections and equipment.
Check the transmitter configuration.
7.3.1 Analog Output
An analog output problem can appear as one of the following:
No output or very low output – There is no transmitter output or the output remains low despite
changes in the process.
High output – Transmitter output remains high despite changes in the process.
Erratic output – Transmitter output varies when process does not.
Sluggish response – Transmitter seems to respond to process changes very slowly.
Often an analog output problem is caused by incorrect transmitter configuration or by something external
to the transmitter.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens 7MF4233 series and is the answer not in the manual?

Siemens 7MF4233 series Specifications

General IconGeneral
BrandSiemens
Model7MF4233 series
CategoryTransmitter
LanguageEnglish

Related product manuals