Calibration and Maintenance UMSITRPSD3-1
November 2005
7-10
Device Parameters After Replacing Electronics Module or Measuring Cell
Parameter Description Stored Not Stored
(2)
Configuration
Dependent (1)
Damping x
AO alarm type x
Alarm LRV/URV x
Saturation alarm activation x
Transmission functions x
Square root switch point x
Pressure units x
Polling address x
Tag x
Mechanical construction:
electronics/sensor
x
Min/max pointer x
Zero calibration x
Upper/lower sensor calibration points x
Service interval x
Minimum span x
Measurement limits
Sensor limits
Explosion protection x
Sensor serial number x
Working hour meter, sensor x
Working hour meter, electronics x
Firmware/Software revision x
Min/max pointer, electronics x
Calibration interval, electronics x
Saturation alarm duration x
Local display unit x
Current calibration
Message x
Descriptor x
Date x
Final assembly number x
Bargraph activation x
Write protection mode x
Hardware revision x
Device revision x
Unit tracking x
Notes:
(1) Column specifies sensor data that remains after changing electronics or sensor.
(2) Dependent on electronics module version. Check the parameters and configure as needed.