EasyManua.ls Logo

Siemens ADVIA 1800 - Analytical Parameter Audit Trail

Siemens ADVIA 1800
222 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
114 ADVIA 1800 Operator’s Guide
Analytical Parameter Audit Trail
ADVIA 1800 software provides an Audit Trail to track user and APTool changes to any
field in the Analytical Parameter window for all methods. The Audit Trail excludes fields
from the Analytical Parameters (Serum) and the ISE Parameter Setting windows.
To view, print, or save to file the Audit Trail:
1. From the menu panel, select Maintenance.
2. From the menu, select Analytical Parameter Audit Trail.
In tabular form, the Audit Trail provides
the name of the screen where the change was initiated
the date and time of the change
the User Code
the test name
Version of the Chemistry Method before the change
Version of the Chemistry Method after the change
Analy. Cond. No.
Sub Param. #
the previous parameter value
the new parameter value
Parameter
The Audit Trail can be purged when the user is logged in as the tech_manager.

Table of Contents