72 ADVIA 1800 Operator’s Guide
1. After the power is applied and the Windows operating system is loaded, the ADVIA
Chemistry system Startup window displays.
2. Turn on the optional universal rack handler:
For the rack handler, set the control panel Standby/On switch to | (ON).
The ready indicator is green.
For the universal rack handler, set the display panel Ready/Standby switch to
READY.
The power indicator is green.
3. At the analyzer power panel, set the Operate/Standby switch (
1) to Operate.
Figure 2-1. Power panel at Startup
The power indicator (2) is on, and the Start (4) and Ready (5) indicators start flashing
when the communication to the PC is established.
4. At the Startup window, do the following:
a. In the Please enter password box, enter the user password.
User passwords assigned in the User Code Settings window are not case sensitive.
b. Select New Start or Re-start, then select
OK.
After a few minutes, the Menu Panel and the Operation Panel open.
5. When the Start and the Ready indicators are off, and the Initialize button on the
Operation Panel activates (turns black), select
Initialize.
6. At the Windows desktop, double-select the rack handler icon.
7. Log on as
supervisor or tech_manager, if required.
Performing a system-assisted startup (Start set)
1. Verify that sufficient system reagents and wash solutions are available for the startup
you want to run.
2. At the Menu Panel, select
Maint. then select System Startup/Shutdown Settings.