190 ADVIA 1800 Operator’s Guide
During operation system settings area
1. In the Realtime monitor area, select from the following formats for routine analysis:
Standard Conc. (ABS) Conc. (ABS – RB)
Conc. Detail
2. In the On-line field, select Do or Not do.
3. In the ADVIA QC Transfer area, select
Yes or No.
4. In the syngo Transfer field, select
Yes or No.
Pre-operation settings
NOTE: You can change these options only before you initialize the system. After the
system is initialized, the options are not available on the screen
.
Specify if these system components are operational or cancelled:
• Chemistry
• ISE
• Unused
• Labo.Auto.Sys./Rack Handler
Clot Detection settings
1. In the Clot detect area, select Avail. to use the clot detection feature or N.A. to
disable it.
2. In the Clot sensitivity area, select
Low, Middle, or High to select the sensitivity of the
clot detection sensor.
Setting up the ISE
This section describes the steps required if the present configuration for the electrolyte
analyzer (ISE) is ever lost or requires modification.
Activating the ISE
1. Set the ELA switch in the power supply distribution chassis panel (on the back of the
analyzer) to the
ON position.
2. At the Systems Specifications Settings window, set the Electrolyte option to
Avail.
3. At the System Monitor window, set the electrolyte option to
Operate.