ISE Operation window
Use this window to perform various ISE tasks.
There is a display area at the top of the window where you can view the running status.
Perform any of the following tasks at this window:
buffer prime
calibration
wash electrode
CV check
interval check
selectivity check
final operation
manual operation
ISE line wash
Dilution bowl drain
initialize
batch print
periodic wash
enter information about each electrode
ISE Monitor window
Perform any of the following tasks at this window:
monitor and verify calibration data as a function of time, using history graphs
monitor serum or urine calibrations
monitor calibration trace data
display data from past calibrations
*monitor selectivity check data
print the results
delete calibration trace data
change y-axis scale
* The selectivity check must be performed under the supervision of authorized Siemens service
personnel. Please call your local technical support provider or distributor.
Manual Operation window
Perform any of the following tasks at this window:
operating a unit
activating or deactivate a unit
checking unit position values
BIOHAZARD
Wear personal protective equipment. Use universal precautions.
Manual operation of units may be necessary when performing maintenance and
troubleshooting tasks.
The window contains a graphical image of the analyzer which depicts the operation you
are performing.