Maintenance 135
12. Only if you replaced the lamp, select Regist Data, then select OK in the Registration
window.
13. If not, proceed to step 14.
NOTE: The system uses the data from the lamp energy data registration as the
comparison standard for the next calculation of the attenuation ratio. The displayed
attenuations for all the reference values are set to 100%, indicating no attenuation.
14. At the Menu Panel, select
System(s), then select Screen Print to print the window
contents.
15. Save the printout with your laboratory maintenance records.
16. Exit the Lamp Energy Monitor window, then select
Initialize to switch the system
from the WAIT state to the READY state.
17. Run cell blank measurement.
Reading lamp energy data
1. Execute the lamp energy check and acquire the data.
• A graph displays at the window.
• The lamp energy check data displays to the left of the graph.
2. Select a wavelength for which to display data.
3. Display the voltage or the AD value.
4. (Optional) Change the vertical scale of the graph.
Measuring cuvette blanks
NOTE: When performing weekly maintenance, be sure to perform the procedures in the
following sequence:
1. Weekly wash
2. Lamp energy check
3. Cuvette blank measurement (this procedure)
Materials required:
• No materials required
Time: 20 minutes
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
Reaction cuvettes cause changes in absorption with use. After the weekly wash, perform
the cuvette blank measurement to determine the change. The cuvette blank is only run
weekly, even if your lab runs the Weekly Wash as a daily procedure.
1. At the Menu Panel, select
Maint, then select User Maintenance.
2. In the Cell blank meas. check area, select
Start CB.
The measured cuvette blank values for all RRV cuvettes and a list of abnormal cells are
available in the real-time monitor in approximately 15 minutes.
3. Select
Yes to save the data.