106 ADVIA 1800 Operator’s Guide
Reviewing the calibration
NOTE: For detailed information on how to use the View Calibration Curve and ISE
Monitor windows, refer to the online Operator’s Guide.
Use the View Calibration Curve window to review the calibration of photometric
methods.
Use the ISE Monitor window to review the calibration of electrolyte methods.
Using the Calibration/RBL History window
1. At the Menu Panel, select Calibration, then select Calibration/RBL History.
2. Select the calibration data to display by selecting Latest or Select:
Latest—Displays the most recent calibration data for each method on the system.
SelectActivates the Test Name box to allow a more detailed test selection.
3. If the Select option is chosen, then select one of the 3 Test Name options:
Photometric—Displays all calibration history data for all photometric methods.
ISE TestsDisplays all calibration history data for all ISE methods.
OtherActivates the test list box to allow specific test selection (highlight each
test to display).
4. Specify a date range, to display data for calibrations performed on specific days for
the tests selected:
a. Type the
Start Date (yyyymmdd) of the range (if other than the default, which is
the current date).
b. Type the
End Date (yyyymmdd) of the range (if other than the default, which is
the current date).
c. Select
Search to display the calibration data for the tests selected for the
specified date range.
NOTE: Select and drag the column heading borders to resize the columns in the
calibration data table.
5. To clear the data for the method selected, select
Clear.
6. When prompted, select
Yes.