EasyManua.ls Logo

Siemens ADVIA 1800 - Set up of Daily Washes

Siemens ADVIA 1800
222 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
System Overview 77
3. After replacing the reagent(s), you must establish the correct on-system
stability.
a. Minimize the Reagent Inventory window.
b. At the Menu Panel, select
Reagent, then select Reagent Container
Settings
.
c. At the Reagent Container Settings window, locate the replaced
nonbarcoded reagent.
d. Enter the correct reagent Exp. date in the form YYYYMMDD.
e. Select the
O (Open) button to open the replaced reagent container and
initialize the Days Remaining counter.
NOTE: If you replace only 1 of the reagents of a 2-reagent method and the
replaced reagent has a new lot number, the system alerts you that a reagent
pair does not exist. You should change both reagents at the same time. If
the 2 reagents that you replaced have new lot numbers, you must re-
calibrate the new set of reagents before you continue to run samples.
(1) At the Menu Panel, select Reagent, then select Reagent Inventory.
(2) Execute a Barcode Scan at the Reagent Inventory window.
(3) Evaluate calibration status.
4. Perform a start-up wash.
Set up of daily washes
WASH 3 or Startup wash, 26 minutes
CTT position 51
DI water (tube)
RTT1 position 56
DI water (70-mL container)
RTT2 position 56 DI water (70-mL container)
WASH 2 or Shutdown wash, 38 minutes
CTT position 15 ISE detergent (cup)
CTT position 16
DI water (tube)
CTT position 49
10% cuvette wash (tube)
CTT position 50
DI water (tube)
RTT1 position 55 10% cuvette wash (70-mL container)
RTT1 position 56
DI water (70-mL container)
RTT2 position 55 10% cuvette wash (70-mL container)
RTT2 position 56
DI water (70-mL container)
Weekly wash set up
Same as WASH 2, except replace the 10% cuvette wash with 5% reagent
probe wash in both the CTT and RTT positions.

Table of Contents

Related product manuals