172 ADVIA 1800 Operator’s Guide
Washing all the ISE lines
Materials required:
2 clean, empty buffer bottles
Probe Wash 3 solution
Dummy electrode
Phillips head screwdriver
Time: 15 minutes
Analyzer mode: Manual operation
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
1. At the Menu Panel, select Maint., then select ISE Operation.
2. In the Period.wash area, select
OFF, then select Set.
3. Open the front doors and replace the buffer solution (
1) with another buffer bottle
containing 500 mL of deionized water.
Figurer 5-36. Location of ISE Buffer bottle
4. Loosen the thumb screw and lift the ISE cover.
5. Disconnect the electrode connectors.
6. Remove the thumbscrew (
1) to release the plate that secures the electrodes and the
block containing the electrode.