System Overview 89
Monitoring samples loaded on the optional universal rack handler
1. At the universal rack handler Operation Panel window, select View on the menu bar,
then select
Rack Handler Operation Monitor or press F8.
Perform any of the following tasks at the Rack Handler Operation Monitor window:
follow the progress of each rack through the rack handler
locate racks that contain sample tubes with non-read labels
find a specific rack or tube currently on the rack handler
skip un-sampled racks
clear rack ID numbers where non-read ID samples occurs
edit non-read tube IDs
Perform the following tasks at the universal rack handler:
enter a sample ID for non-read labels at the Edit Non-Read Tube ID window
obtain a summary of the current rack handler activity at the Data Monitor window
view and print all the racks/samples processed by the rack handler on a day-by-day
basis at the Review Data window
2. Use the Review/Edit window on the ADVIA software to determine when results are
available.
Running an interrupt (STAT) sample
Running an interrupt sample when sampling is in progress (START)-SMP
LOAD NG
1. Download or create the sample workorder and select
Interr. (interrupt) option on the
workorder.
NOTE: You can also select the priority using the Cup/Tube Assign window if the
sample is already on the sample tray and already has a workorder.
2. At the Operation Panel, select Pause to halt sampling.
3. Load the new sample on the sample tray (STT).
4. At the Operation Panel, select
Start to resume sampling.
Running an interrupt sample when sampling has stopped (STOP)-SMP
LOAD OK
1. Select Interr. (interrupt) option on the sample workorder.
2. Load the sample on the sample tray (STT).
NOTE: Priority can be set using workorder Interr. designation and the Cup/Tube
Assign or Sample confirmation window that displays after reading the barcoded
samples.
3. To begin sampling, at the Operation Panel, select
Start.