Appendix 211
Item Description
• System is for indoor use at an altitude of up to 2000 meters with a
pollution degree of 2.
• Maximum relative humidity allowable with system operating is 40% to
70% with no condensation.
• Heat output:
50 Hz 60 Hz
Power off mode:
400 W 350 W
1365 Btu/hr 1024 Btu/hr
344 kCal/hr 300 kCal/hr
Ready mode:
1260 W 1030 W
4299 Btu/hr 3023 Btu/hr
1084 kCal/hr 886 kcal/hr
Auto mode:
1480 W 1540 W
5050 Btu/hr 4518 Btu/hr
1273 kCal/hr 1324 kCal/hr
Water requirements
The system is connected directly to a pressurized water source using
Type 1 CLSI/CAP
or ISO 3696 water.
At a minimum, the water supply must meet the following parameters to
ensure proper operation, as per the Site Survey for each ADVIA
Chemistry system:
• Resistivity > 10.0 Megaohm-cm @ 25°C
• Bacteria < 10 cfu/mL
• Silica < 0.05 mg/L
• Total Organic Carbon < 200 µg/L
• Total Dissolved Solids < 1.0 mg/L
In addition, the following specifications should be met to ensure proper
operation:
• Temperature of water entering the system is between 10°C and 30°C
• Dissolved Oxygen ≤ 7.0 mg/L
Direct plumbing deionized water pressure: 9.6 - 96 kPa (1.4 - 14.2 psi)
Drain requirements
Minimum of 40 liters (10.6 gallons) per hour
If the local laboratory practices and/or applicable environmental
regulations prohibit the inclusion of concentrated waste into the
laboratory's drain, an optional concentrated waste bottle must be ordered.
Transportation and Storage
conditions
Temperature -29 to +60°C
Fragile, handle with care. Keep dry, do not stack, do not overturn.