Prerequisites 1 - 7
Siemens AG SPR2-130.061.02 Page 7 of 10 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
Covers 1
Basic system 1
Back cover
Removing the back cover
• Put on the foot brake.
• Remove the screws from the back cover.
• Pull the cover back approx. 25 cm.
• Unscrew the protective conductor from the cover.
• Pull the cover completely off and tilt it downward.
• Lift up the cover and raise both lateral metal brackets from the guide rails.
Installing the back cover
• Fit the lateral metal brackets back into the guide rail.
• Remount the protective conductor to the back cover.
• Lift the cover and push it forward. Be careful with the EMC spring contact.
• Replace the cover and tighten the screws.
Monitor trolley 1
Back covers
Removing the covers
• Remove the upper back cover plate of the log book compartment.
• Remove the protective conductor.
• Remove the lateral Allen screws from the lower back cover.
• Remove the protective conductor from the cover.
Installing the covers
• Begin with the lower back cover.
• Remount the protective conductor for the lower cover.
• Install the cover and tighten the screws.
• Remount the protective conductor for the Logbook cover.
• Install the cover and tighten the screws.
Electrical voltage!
See Chapter 1, Safety Information.
Disconnect the line voltage plug prior to removing covers.
WARNING