Adjustments / Programs 5 - 15
Siemens AG SPR2-130.061.02 Page 15 of 30 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
Collimator displays on the monitor 5
Prerequisites 5
• Camera rotation, collimators for X-Iris and slot diaphragm have been set correctly.
Collimator display, X-Iris 5
• Connect the service PC to the serial interface of the SIREMOBIL.
• Start the service program.
• Select the "Adjustment, Calibrations..." menu.
• Respond to the query with "Yes".
• Select "Screen Display Collimator" in the "Function Groups:" combo box.
Start adjustment
• In the "Available Functions:" window select "Init" and click on "Execute". The X-Iris is au-
tomatically opened to full format and the slot diaphragm is opened to maximum aperture.
Full format, X-Iris open
• Zoom must be disabled.
• Release fluoro and wait until generator has performed automatic adjustment.
• Switch fluoro off. The X-Iris is displayed on the monitor.
• Use the keys for opening and closing the X-Iris to set the diameter of the superimposed
circle to the size of the X-iris.
• In the "Available Functions:" window select "Open Pos - II. Full Format" and click on "Ex-
ecute".
• Close the status window by clicking on the OK button.
• For the query "Do you want to reset the unit", click on "No". The X-Iris is automatically
closed to the smallest format.
Full format, X-Iris closed
• Briefly release fluoro. The X-Iris is displayed on the monitor.
• Use the keys for opening and closing the X-Iris to set the diameter of the superimposed
circle to the size of the X-Iris.
• In the "Available Functions:" window select "Close Pos - II. Full Format" and click on "Ex-
ecute".
• Close the status window by clicking on the OK button.
• For the query "Do you want to reset the unit", click on "No".
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
WARNING