3 - 1
Siemens AG SPR2-130.061.02 Page 1 of 2 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
Checking the operating voltages 3
Low voltages 3
Check the operating voltages for the SIREMOBIL Compact against the values in Table 1.
I.I. voltages 3
The voltages E1 / E2 / E3 and A may be taken from the I.I. test protocol and checked or
adjusted against the control test points listed in Table 2.
from test point to test point Voltage Tolerance Potentiometer
M14.4 (0V) M14.5/6 (+5 V) + 5 V +
0.2 V M14. +5 V/Adj.
M14.3 (0V) M14.2 (+15 V) + 15 V +
0.1 V M14. +15 V/Adj.
M14.3 (0V) M14.1 (-15 V) - 15 V +
0.5 V M14. -15 V/Adj.
D3.X8.1 (0V) D3.X8.3 (+26,75 V) +26.75 V +
0.25 V D3.R3
M13.S- (0V) M13.S+ (+13 V) +13 V +
0.2 V M13.TR1
D3.X1.7 (0V) D3.X1.9 (+24 V) + 26.5 V to
+ 29.5 V
n.a. not adjustable
Tab. 1
Electrical voltage!
See Chapter 1, Safety Information.
Do not touch any part of the I.I. mini voltage supply!
Voltage Test point Ground
point
(0V)
Potentiome-
ter for full
format
Potentio-
meter
for zoom
format
Voltage
divider
ratio
E1 UE1 P10 P11 1:1
E2 UE2 P6 P7 1:1
E3 UI 15 P2 P3 1:10000
30 kV, Anode UI 30 P1 P1 1:10000
Tab. 2
WARNING