SIREMOBIL Compact SPR2-130.061.02 Page 12 of 30 Siemens AG
System Manual Rev. 11 09.04 CS PS 24 Medical Solutions
5 - 12 Adjustments / Programs
• Select I.I. survey format.
• During fluoroscopy, adjust the X-Iris by opening and closing it with the X-Iris keys so that
the collimator blades are still visible along the edge of the I.I.
• Stop fluoroscopy.
• Select "2. II. - Full Format Position" on the service PC and click on "Execute".
• Close the status window by clicking on the OK button.
• For the query "Do you want to reset the unit," click "No".
• Switch to zoom format.
• During fluoroscopy, adjust the X-Iris by opening and closing the X-Iris keys so that the
collimator blades are still visible along the edge of the I.I.
• Stop fluoroscopy.
• Select "3. II. - Zoom Format Position" on the service PC and click on "Execute".
• Close the status window by clicking on the OK button.
• For the query "Do you want to reset the unit," click "No".
• Open the X-Iris completely.
• Select "4. CEX Position" on the service PC and click on "Execute".
• Close the status window by clicking on the OK button.
• For the query "Do you want to reset the unit," click "No".
• Close the X-Iris completely.
• Release fluoroscopy. The X-Iris must close between 4 - 4.5 cm.
• Stop fluoroscopy.
• Select "5. CLOSED POSITION" on the service PC and click on "Execute".
• Close the status window by clicking on the OK button.
• For the query "Do you want to reset the unit," click "NO".
• Select "8. Init Off & Store Positions" on the service PC and click on "Execute".
• Close the status window by clicking on the OK button.
• For the query "Do you want to reset the unit," click "Yes".
• Switch the SIREMOBIL Compact OFF and ON.
• Check the X-Iris again in survey format and zoom format. The collimator blades must still
be visible on the edge of the image. If necessary, repeat the adjustment.
• Stop fluoroscopy.
Concluding steps 5
• Return to the main menu on the service PC with the ESC key.
• Select the Data, Backup... menu.
• Select Parameter in the backup window.
• Click on "Backup". At "Remarks for Parameter" enter "X-Iris set".