SIREMOBIL Compact SPR2-130.061.02 Page 12 of 30 Siemens AG
System Manual Rev. 11 09.04 CS PS 24 Medical Solutions
5 - 12 Adjustments / Programs
Select I.I. survey format.
During fluoroscopy, adjust the X-Iris by opening and closing it with the X-Iris keys so that
the collimator blades are still visible along the edge of the I.I.
Stop fluoroscopy.
Select "2. II. - Full Format Position" on the service PC and click on "Execute".
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit," click "No".
Switch to zoom format.
During fluoroscopy, adjust the X-Iris by opening and closing the X-Iris keys so that the
collimator blades are still visible along the edge of the I.I.
Stop fluoroscopy.
Select "3. II. - Zoom Format Position" on the service PC and click on "Execute".
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit," click "No".
Open the X-Iris completely.
Select "4. CEX Position" on the service PC and click on "Execute".
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit," click "No".
Close the X-Iris completely.
Release fluoroscopy. The X-Iris must close between 4 - 4.5 cm.
Stop fluoroscopy.
Select "5. CLOSED POSITION" on the service PC and click on "Execute".
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit," click "NO".
Select "8. Init Off & Store Positions" on the service PC and click on "Execute".
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit," click "Yes".
Switch the SIREMOBIL Compact OFF and ON.
Check the X-Iris again in survey format and zoom format. The collimator blades must still
be visible on the edge of the image. If necessary, repeat the adjustment.
Stop fluoroscopy.
Concluding steps 5
Return to the main menu on the service PC with the ESC key.
Select the Data, Backup... menu.
Select Parameter in the backup window.
Click on "Backup". At "Remarks for Parameter" enter "X-Iris set".