Siemens SIREMOBIL Compact L - Correction of the 0° Position of the Slot Diaphragm Display

Siemens SIREMOBIL Compact L
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Adjustments / Programs 5 - 19
Siemens AG SPR2-130.061.02 Page 19 of 30 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
Correction of the 0° position of the slot diaphragm display 5
Select the "Adjustments," "Parameter" menu at the service PC.
Select "Screen display Slot diaphragm" in the Combo box.
Click "Get from Unit".
Select "Rotation 0 degree position".
The parameter in the "Value Actual" field can be changed according to the minimum and
maximum values.
Then click "Set Value".
Click "Put to Unit".
After the SIREMOBIL Compact has booted, compare the 0° position of the slot dia-
phragm blades to the display on the monitor. If necessary, repeat the programming.
Click "Save to file" .
Leave the service program with "Logoff" and "Quit" to save the parameters as a new file
to disk.
Correct only if the slot diaphragms are shown at an angle of rota-
tion different from that displayed on the monitor. A correction of
+
10° is possible. Camera rotation and slot diaphragm have to be
set correctly.
The service PC is connected to the serial interface of the
SIREMOBIL Compact.
The service program is started.
NOTE