Adjustments / Programs 5 - 19
Siemens AG SPR2-130.061.02 Page 19 of 30 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
Correction of the 0° position of the slot diaphragm display 5
• Select the "Adjustments," "Parameter" menu at the service PC.
• Select "Screen display Slot diaphragm" in the Combo box.
• Click "Get from Unit".
• Select "Rotation 0 degree position".
• The parameter in the "Value Actual" field can be changed according to the minimum and
maximum values.
• Then click "Set Value".
• Click "Put to Unit".
• After the SIREMOBIL Compact has booted, compare the 0° position of the slot dia-
phragm blades to the display on the monitor. If necessary, repeat the programming.
• Click "Save to file" .
• Leave the service program with "Logoff" and "Quit" to save the parameters as a new file
to disk.
Correct only if the slot diaphragms are shown at an angle of rota-
tion different from that displayed on the monitor. A correction of
+
10° is possible. Camera rotation and slot diaphragm have to be
set correctly.
The service PC is connected to the serial interface of the
SIREMOBIL Compact.
The service program is started.
NOTE