Replacing boards / replacing components 6 - 17
Siemens AG SPR2-130.061.02 Page 17 of 28 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
MEMOSKOP C-E 100 6
Replace the MEMOSKOP C-E 100.
The attachment screws are accessible on the underside of the storage device.
Check the software status of the MEMOSKOP. Download the system software status to
the MEMOSKOP if necessary.
Use the MEMOSKOP Service Software to import the language file.
Set the MEMOSKOP C-E 100 to the correct video frequency and configure it for the
number of existing monitors.
Refer to Chapter 5 "Configuring the video frequency " and "MEMOSKOP - Configuration
for the number of existing monitors".
Configure the following parameters in Technical Setup for MEMOSKOP:
Check the software status of the newly delivered MEMOSKOP.
If the software status of the MEMOSKOP differs from that of the
system, the system software status must be downloaded to the
MEMOSKOP. The MEMOSKOP software status is displayed in the
2nd line of the User Setup. The first number describes the
MEMOSKOP type (2=CE; 3=CE-100/ C/ C & MOD; 4=C-SUB/ C-SUB
& MOD).
The image storage MEMOSKOP C-E 100 is a new image storage
that works only with the correct number of images. For this rea-
son, if the image storage was set to it's default values
(Menu "Diagnostics," "Load Factory Defaults"), the parameter
"Maximum image numbers" must be set to 100 images.
Read and record the parameters for MEMOSKOP C-E 100 if you
can select User setup. After replacing the MEMOSKOP, configure
the MEMOSKOP C-E 100 according to the parameters noted.
Number of monitors Configure 1 for SIREMOBIL Compact with 1 monitor.
Configure 2 for SIREMOBIL Compact with 2 monitors.
Hardcopy Check whether 5000 is programmed in "4 Max Hard-
copy time". If not, program this value.
Edge 1 coef set Check whether 3 is programmed, If not, program this
value.
Edge 2 coef set Check whether 8 is programmed, If not, program this
value.
Motion detection setup Check whether the following are programmed "MD1:
Threshold" 100 and for "MD2: Threshold" 120. If not,
program these values.
Date / Time format Set the country-specific format.
Keyboard Set to the keyboard layout being used.
Language Set for language of the country.