SIREMOBIL Compact SPR2-130.061.02 Page 16 of 28 Siemens AG
System Manual Rev. 11 09.04 CS PS 24 Medical Solutions
6 - 16 Replacing boards / replacing components
• Configure the following parameters in User setup for MEMOSKOP:
• Perform the IQ quick test.
Default Edge Configure 1. Program a different Edge enhancement
value only upon customer request.
Monitor split Configure Yes.
Monitor text Configure Yes.
Disable dose rate selection in
user setup
Set to "No" if equipped with an area dose product mea-
surement device. Menu "Storage of Dose on Disk" and
Menu "Dose on Hardcopy output in User Setup are visi-
ble. Otherwise, set to "Yes".
Unit serial number Program the serial number of the SIREMOBIL Compact.
Do not enter the first zero.
Product name SIREMOBIL Compact (default). If wrong, configure
"SIREMOBIL Compact".
Default program changeable Default: Yes.
Depending on country specific regulations, configure
Yes or No.
Software backward compatible Configure No.
Operating mode Parameters Upon customer request, change the parameters pro-
grammed for the organ program.
VCR output Configure to Monitor A.
Hardcopy Release For SIREMOBIL Compact with 1 monitor:
Configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
Configure to monitor B.
Text monitor For SIREMOBIL Compact with 1 monitor:
Configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
Configure to both monitors.
Split Screen Configure horizontally. Configure vertically only upon
customer request.
Display of time Configure Yes.
Set Date Time Set the current date and time.
Hospital Name Program the hospital name.
Storage Dose on Disk If the area dose product meter is present, (Diamentor)
configure Yes. Otherwise, configure No.
Dose on hard copy output If the area dose product meter is present, (Diamentor)
configure Yes. Otherwise configure No.