Adjustments / Programs 5 - 11
Siemens AG SPR2-130.061.02 Page 11 of 30 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
Collimator, X-Iris 5
Prerequisites 5
• Attach the center cross in the center of the image intensifier input screen.
• Connect the service PC to the serial interface on board D1.
• Start the service program.
Adjustments 5
• Select the "Adjustment, Calibration" menu.
• Answer the query with "yes".
• Select "X-Iris" in the "Function Groups" combo box.
• In the window "Available Functions:" select "1. Init (Limits Off / Autosearch)" and click on
Execute.
• Close the status window by clicking on the OK button.
• For the query "Do you want to reset the unit," click "No".
• The X-Iris can now be completely opened and closed with the X-Iris keys.
Because of the mechanical hysteresis of the semi-transparent
slot diaphragm, open the collimator blades to maximum aperture,
prior to adjusting the radiation iris.
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
NOTE
WARNING