Adjustments / Programs 5 - 23
Siemens AG SPR2-130.061.02 Page 23 of 30 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
• Start the service program.
• Select the Adjustment, Calibrations menu.
• Click "Yes" to answer the query.
• In the Combo box "Function Group:," select "Diamentor".
• In the window "Available Functions", select the line "Start Diamentor Test" and click on
"Execute".
• Close the status window by clicking on the OK button.
• Respond the query "Do you want to reset the unit" by clicking "No".
• Press the -0- key (reset key for fluoroscopic time).
• The tested Qt*Kg value is displayed on the monitor of the SIREMOBIL Compact.
• This value must match the programmed CONSTANT (Qt*Kg) value.
Tolerance: +
5%.
• By pressing the -0- key (reset key for fluoroscopic time) the test is repeated and is dis-
played again on the monitor.
• The potentiometer CAL on the electronic system of the area dose product measurement
device can be used to adjust the test value to the calculated and programmed Qt*Kg val-
ue.
• After the adjustment, select "Stop Diamentor Test" on the service PC and click on "Exe-
cute".
• Close the status window by clicking on the OK button.
• Respond the query "Do you want to reset the unit" by clicking "No".
• The test is ended.
• Check the accuracy of the area dose product measurement device. Refer to the next
section.
Checking the accuracy of the area dose product measurement
device
5
Prerequisites 5
The calibration of the area dose product measurement device has already been per-
formed.
Preparations 5
• Attach and center the center cross on the I.I. (Fig. 6).
• Select survey format.
• Select KV mA curve plateau HC1.
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
WARNING