SIREMOBIL Compact SPR2-130.061.02 Page 10 of 28 Siemens AG
System Manual Rev. 11 09.04 CS PS 24 Medical Solutions
6 - 10 Replacing boards / replacing components
Replacing the Compact optics. 6
Removing the Compact optics 6
Turn the C-arm so that the I.I. is at the top.
Loosen and remove the I.I. cover.
Rotate the camera in the "0" position (6/Fig. 6).
Fig. 5 Fig. 6
Fig. 7 Fig. 8
Electrical voltage!
See Chapter 1, Safety Information.
Prior to removing the mini voltage supply, the system must be
switched off for at least 3 minutes until the high voltage in the
system and in the I.I. mini voltage supply dissipates.