SIREMOBIL Compact SPR2-130.061.02 Page 16 of 30 Siemens AG
System Manual Rev. 11 09.04 CS PS 24 Medical Solutions
5 - 16 Adjustments / Programs
• The X-Iris remains in the minimum position.
• Zoom format is automatically selected.
Zoom format, X-Iris closed
• Briefly release fluoro. The X-Iris is displayed on the monitor.
• Use the keys for opening and closing the X-Iris to set the diameter of the superimposed
circle to the size of the X-Iris.
• In the "Available Functions:" window select "Close Pos - II. Zoom Format" and click on
"Execute".
• Close the status window by clicking on the OK button.
• For the query "Do you want to reset the unit" click on "No".
• The X-Iris is opened automatically.
• Zoom format remains selected.
Zoom format, X-Iris open
• Briefly release fluoro. The X-Iris is displayed on the monitor.
• Use the keys for opening and closing the X-Iris to set the diameter of the superimposed
circle to the size of the X-Iris.
• In the "Available Functions:" window select "Open Pos - II. Zoom Format and click on
"Execute".
• Close the status window by clicking on the OK button.
• For the query "Do you want to reset the unit", click on "No".
Save values and end adjustment
• In the "Available Functions:" window select "INIT OFF & STORE VALUES" and click on
"Execute".
• Close the status window by clicking on the OK button.
• For the query "Do you want to reset the unit", click on "Yes".
Concluding tasks 5
Before ending the service program, save the new parameter values to disk
(Service PC, Data menu, Backup).