SIREMOBIL Compact SPR2-130.061.02 Page 2 of 2 Siemens AG
System Manual Rev. 11 09.04 CS PS 24 Medical Solutions
7 - 2 Brake force / Brake torque
Lubricating the vertical column 7
• The spindle in the vertical column should be lubricated during annual maintenance.
• Raise the vertical column approximately 50 cm. A drilled hole is visible in the column.
• Fill this hole with approximately 2 cm
3
of oil.
Oil to be used:
Special purpose oil (Optimol GmbH; Viscogen KL300; 40 g),
Item number 73 95 353 RH090.