Adjustments / Programs 5 - 7
Siemens AG SPR2-130.061.02 Page 7 of 30 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
Setting the dose rate for survey format 5
Select the I.I. survey format (refer to the I.I. ZOOM STATUS).
Center the dose measurement chamber on the I.I. input screen and attach.
Use the key to select the dose rate (LOW / MID / HIGH, shown in the "DOSERATE"
line).
Apply fluoro briefly until the generator finishes regulating the automatic fluoro adjust-
ment.
Select KV STOP at the system operating console.
Switch on fluoro.
Set the dose rate with the [+kV] and [-kV] keys.
(Dose rate * grid correction factor (* measurement chamber correction factor - if need-
ed)).
The resulting KV value should be between 70 to 80kV . If necessary, change the prefiltra-
tion and set the dose rate again using keys [+kV] or [-kV]. Refer to the table.
Switch off fluoro.
Remove the dose measurement chamber from the I.I. input screen.
Switch on fluoro.
Press the [>0<] at the system operating console. This starts the automatic adjustment
procedure for the TV- iris collimator.
Wait until the message "done" appears in the "AUTO REGULATION" line.
The "BRIGHTNESS DEVIATION" value is shown in the "BRIGHTNESS DEVIATION"
line and should be 0 +
3.
Switch off fluoro.
Setting the dose rate for zoom format 5
refer also to notice on previous page *.
KV-STOP remains selected.
Center the dose measurement chamber on the I.I. input screen and attach.
Select the I.I. zoom format (refer to "I.I. ZOOM STATUS"). The KV values remain the
same. The mA values, however, are automatically increased by a factor of 1.41.
Use the [mA+] and [mA-] keys to set the dose rate for the zoom format
(Dose rate *grid correction factor (* measurement chamber correction factor - if need-
ed)).
Press the [>0<] key at the system operating console. This starts the automatic adjust-
ment procedure for the TV-iris collimator.
Wait until the message "done" appears in the "AUTO REGULATION" line.
The "BRIGHTNESS DEVIATION" value is shown in the "BRIGHTNESS DEVIATION"
line and should be 0 +
3, refer also to notice on previous page *.
Switch off fluoro.