EasyManua.ls Logo

Siemens SIREMOBIL Compact L - Camera rotation

Siemens SIREMOBIL Compact L
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SIREMOBIL Compact SPR2-130.061.02 Page 20 of 30 Siemens AG
System Manual Rev. 11 09.04 CS PS 24 Medical Solutions
5 - 20 Adjustments / Programs
Camera rotation 5
Fig. 3
Fig. 4 Fig. 5
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
The keys for horizontal and vertical image reversal must be off
prior to adjusting camera rotation. Camera rotation should be set
in the 0° position. Fig. 3, Fig. 4 and Fig. 5 show the 0° position.
With the right camera rotation key (the arrow on the key is point-
ing clockwise), the marker on the monitor will rotate clockwise.
The image contents will rotate counterclockwise.
With the left camera rotation key (the arrow on the key is pointing
counterclockwise), the marker superimposed on the monitor will
rotate counterclockwise. The image contents will rotate clock-
wise.
Patient right position
Patient left position
WARNING
NOTE
NOTE