6 - 6 Replacing boards / replacing components
• Check the setting of the laser targeting device and, if necessary, adjust it according to
the Adjustment Instructions for the Laser Targeting Device, RXR2-130.032.01.
• Re-attach the SIREPHOS cover.
• Seal the SIREPHOS cover using sealing compound. See Chapter 1 "Covers".
Replacing the I.I. 6
Checking the temperature indicator 6
• Prior to installing the I.I., check the temperature indicator:
ê If the inner square field of the indicator is white, the I.I. did not exceed the tem-
perature range. Remove the temperature indicator.
ê If the indicator is discolored (inner field is black), proceed according to the
IQ document RXD-000.038.01
.